Rabu, 15 Juni 2011

High power monopolar and bipolar SURTRON 300 HP

SURTRON 300 HP

High power monopolar and bipolar

Can be used to execute monopolar and bipolar surgery in the many fields of application where it is requested high precision and reliability



Can be used to execute monopolar and bipolar surgery in the many fields of application where it is requested high precision and reliability

SURTRON® HP can be used to execute monopolar and bipolar surgery in the many fields of application where it is requested high precision and reliability.
TWIN OPERATIVE ACTIVITY
• Two output channels for monopolar cutting
• Two independent setting and output channels for the most frequent monopolar coagulation

CONTROLLED CUTTING PROCEDURES
• Enhanced cutting current for eschar free incision
• Cutting current with selectable coagulating property for bleeding controlled dissection
• Low capacitive coupling by slow BLEND current for endoscopic procedures

TWIN CONTROLLED COAGULATIVE PROCEDURES
• SPEEDY* coagulation and DEEP**coagulation available to two operators
• SPRAY coagulation available by hand switch or foot switch controls

WIDE BIPOLAR FEATURES

• Bipolar cutting current with selectable bleeding control
• Slow BLEND cutting current
• Separately activable Automatic Start/Stop coagulation (Detection system of tissue impedance)

SURGEON AND PATIENT TOTAL SAFETY
• Acceptability of contact impedance between patientbody and return electrode continuously watched by Skin Plate Electronic Control (SPEC)
• H.F. Leakage currents continuously monitored
• Mains fluctuation automatically compensated before to start the delivery of HF power

PERSONALIZED PROCEDURES
• Independent choise of coagulative currents and power levels by each of two operators
• Ten personal configurable and immediately recallable procedures

EXTENTION DRIVER AND INTERLOCK SIGNALS
• Drived output connection to external Argon Plasma unit and Smoke Evacuator unit

AUTOMATIC SELF DIAGNOSIS
• Automatic diagnostic control of the actual complete functionality of the electronic unit and the connected accessories with coded report of failures

*SPEEDY coagulation also said Fulgurate or Forced coagulation
**DEEP coagulation also said Pin Point, Dessicate or Soft coagulation
CONTROLS
Patient/Plate circuit monitoring
Output power monitoring
Self check control
H.F. Leakage currents control
Compensation of mains fluctuation

SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Electrical Class: I CF
MDD 93/42/EC Class: II b


SURTRON®HP
300 HP
400 HP
Reference code
10400.801
10400.901
Maximum output power monopolar pure CUT
300 W→ 300 Ω
400 W → 300 Ω
Maximum output power monopolar cut-coag CUT80%
250 W → 300 Ω
300 W→ 300 Ω
Maximum output power monopolar cut-coag CUT60%
200 W → 300 Ω
250 W → 300 Ω
Maximum output power monopolar cut-coag CUT40%
150 W → 300 Ω
200 W → 300 Ω
Maximum output power monopolar cut ENHANCED
200 W → 500 Ω
250 W → 500 Ω
Maximum output power monopolar cut-coag BLEND
200 W → 300 Ω
25 W 0 → 300 Ω
Maximum output power monopolar coag SPEEDY
100 W → 500 Ω
120 W → 500 Ω
Maximum output power monopolar coag DEEP
100 W → 200 Ω
120 W  → 200 Ω
Maximum output power monopolar coag SPRAY
70 W → 2000 Ω
100 W → 2000 Ω
Maximum output power monopolar coag SPRAY ARGON
70 W → 2000 Ω
100 W → 2000 Ω
Maximum output power bipolar CUT
95  W → 150 Ω
95  W → 150 Ω
Maximum output power bipolar cut-coag CUT80%
95  W → 150 Ω
95 W → 150 Ω
Maximum output power bipolar cut-coag CUT60%
95  W → 150 Ω
95  W → 150 Ω
Maximum output power bipolar cut-coag CUT40%
60 W → 150 Ω
60 W → 150 Ω
Maximum output power bipolar cut-coag BLEND
90 W → 150 Ω
90 W → 150 Ω
Maximum output power bipolar coag COAG
70 W → 100 Ω
70 W → 100 Ω
Working frequency Monopolar / Bipolar
425 kHz / 525 kHz
425 kHz / 525 kHz
Patient circuit
F
F
Selectable input voltage
115- 230 Vac
115- 230 Vac
Mains frequency
50 - 60 Hz
50 - 60 Hz
Electrical input power
1000 VA
1000 VA
Weight Kgs
17.5
17.5
Size WxHxD mm
470x150x400
470x150x400