Rabu, 15 Juni 2011

Monopolar and Bipolar Radiosurgical unit with an emission of continuous or cycle timer power SURTRON FLASH 160 HF

SURTRON FLASH 160 HF

Monopolar and Bipolar Radiosurgical unit with an emission of continuous or cycle timer power

Electro‐surgical equipment at radio‐frequency (radio‐surgery).


Electro‐surgical equipment at radio‐frequency (radio‐surgery).

SURTRON® FLASH 160  HF , is a High Frequency RadioSurgical equipment which is suitable to precision monopolar and bipolar surgery and micro-surgery without tissue alterations.
SURTRON® FLASH 160  HF , can supply RF operative powers either for programmable time as short as hundreds of a second.
SURTRON® FLASH 160  HF  through its performances allows pure CUT,  cut-coagulation BLEND, incision with reduced prodution of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, BIPOLAR coagulation and ABLATION. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of job. In the bipolar coagulation modality can be activated a detection system of the tissues impedance with automatic activation and automatic stop when it coagulation is achieved (AUTOSTART – AUTOSTOP).
SURTRON® FLASH 160  HF , allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.
CONTROLS
Patient/Plate circuit monitoring
Output power monitoring
Self check control

SAFETY
EN60601-1 
EN60601-1-2 
EN60601-2-2 
Electrical Class: I CF
MDD 93/42/EC Class: II b


Main Features
SURTRON® FLASH
160 HF
Reference code
10300.201
Max output power CUT (1)
150 W→ 250 Ω
Max output power BLEND (1)
100 W → 200 Ω
Max output power ENHANCED (1)
100 W→ 250 Ω
Max output power FORCED COAG
100 W → 150 Ω
Max output power SOFT COAG
80 W → 100 Ω
Max output power BIPOLAR COAG
60 W→ 100 Ω
Max output power BIPOLAR ABLATION (1)
60 W→ 100 Ω
Working frequency
600 kHz
Working frequency ((1)FUNCTION ONLY)
4.0 MHz
RF power pulse emission time
10 – 990 ms
RF power pulse interval
1 – 9
Patient circuit
F
Selectable input voltage
115 - 230 Vac
Mains frequency
50 - 60 Hz
Electrical input power
350 VA
Weight Kgs
6
Size WxHxD mm
370x144x319



Tidak ada komentar: