SURTRON FLASH 200
Monopolar and bipolar with an emission of continuous or cycle timer power
Supplies RF operative powers either for programmable time as short as hundreds of a second.
Supplies RF operative powers either for programmable time as short as hundreds of a second.
SURTRON® FLASH, can supply RF operative powers either for programmable time as short as hundreds of a second.
SURTRON® FLASH, through its performances allows pure CUT, cut-coagulation BLEND, incision with reduced prodution of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, BIPOLAR coagulation and ABLATION. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of job.
In the bipolar coagulation modality can be activated a detection system of the tissues impedance with automatic activation and automatic stop when it coagulation is achieved (AUTOSTART – AUTOSTOP).
SURTRON® FLASH allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.
CONTROLS
Patient/Plate circuit monitoring
Output power monitoring
Self check control
SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Electrical Class: I CF
MDD 93/42/EC Class: II b
Patient/Plate circuit monitoring
Output power monitoring
Self check control
SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Electrical Class: I CF
MDD 93/42/EC Class: II b
SURTRON® FLASH | 120 | 200 |
Reference code | 10300.101 | 10300.401 |
Max output power CUT | 120 W → 250 Ω | 200 W → 250 Ω |
Max output power BLEND | 90 W → 200 Ω | 120 W → 200 Ω |
Max output power ENHANCED | 90 W → 250 Ω | 120 W → 250 Ω |
Max output power FORCED COAG | 80 W → 150 Ω | 150 W → 150 Ω |
Max output power SOFT COAG | 60 W → 100 Ω | 90 W → 100 Ω |
Max output power BIPOLAR COAG | 60 W → 100 Ω | 80 W → 100 Ω |
Max output power BIPOLAR ABLATION | -- | -- |
Working frequency | 600 kHz | 600 kHz |
RF power pulse emission time | 10 – 990 ms | 10 – 990 ms |
RF power pulse interval | 1 – 9 | 1 – 9 |
Patient circuit | F | F |
Selectable input voltage | 115 – 230 Vac | 115 – 230 Vac |
Mains frequency | 50- 60 Hz | 50 - 60 Hz |
Electrical input power | 300 VA | 350 VA |
Weight Kgs | 6 | 6 |
Size WxHxD mm | 370x144x319 | 370x144x319 |