Rabu, 15 Juni 2011

Monopolar and bipolar with an emission of continuous or cycle timer power SURTRON FLASH 200

SURTRON FLASH 200

Monopolar and bipolar with an emission of continuous or cycle timer power

Supplies RF operative powers either for programmable time as short as hundreds of a second.

Supplies RF operative powers either for programmable time as short as hundreds of a second.

SURTRON® FLASH, can supply RF operative powers either for programmable time as short as hundreds of a second.
SURTRON® FLASH, through its performances allows pure CUT,  cut-coagulation BLEND, incision with reduced prodution of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, BIPOLAR coagulation and ABLATION. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of job.
In the bipolar coagulation modality can be activated a detection system of the tissues impedance with automatic activation and automatic stop when it coagulation is achieved (AUTOSTART – AUTOSTOP).
SURTRON® FLASH allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.
CONTROLS
Patient/Plate circuit monitoring
Output power monitoring
Self check control

SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Electrical Class: I CF
MDD 93/42/EC Class: II b


SURTRON® FLASH
120
200
Reference code
10300.101
10300.401
Max output power CUT
120 W → 250 Ω
200 W → 250 Ω
Max output power BLEND
90 W → 200 Ω
120 W → 200 Ω
Max output power ENHANCED
90 W → 250 Ω
120 W → 250 Ω
Max output power FORCED COAG
80 W → 150 Ω
150 W → 150 Ω
Max output power SOFT COAG
60 W → 100 Ω
90 W → 100 Ω
Max output power BIPOLAR COAG
60 W → 100 Ω
80 W → 100 Ω
Max output power BIPOLAR ABLATION
--
--
Working frequency
600 kHz
600 kHz
RF power pulse emission time
10 – 990 ms
10 – 990 ms
RF power pulse interval
1 – 9
1 – 9
Patient circuit
F
F
Selectable input voltage
115 – 230 Vac
115 – 230 Vac
Mains frequency
50-  60 Hz
50 - 60 Hz
Electrical input power
300 VA
350 VA
Weight Kgs
6
6
Size WxHxD mm
370x144x319
370x144x319



Tidak ada komentar: